A compact miniaturized flow system based on low-temperature co-fired ceramic technology coupled to led mini-photometer for determination of dipyrone in pharmaceutical formulations.

No hay miniatura disponible

Fecha

Título de la revista

ISSN de la revista

Título del volumen

Editor

Resumen

Descripción

In this work, an analytical microsystem based on LTCC (low-temperature co-fired ceramic) technology with monolithic incorporation of an optical flow cell for determination of dipyrone in pharmaceuticals is described. The detection system is based on the formation of a blue chromophore between dipyrone and Fe(III) photometrically monitored at 630 nm using a lab-made LED mini-photometer constructed with a light emitting diode as a radiation source and a Si photodiode as a detector. The lab-made mini-photometer elaborated presented a good performance in regard to high signal/noise ratio, low drift and good sensitivity. The analytical curve was linear in the dipyrone concentration range from 1.0 × 10-4 to 3.5 × 10-3 mol L-1, limit of detection of 4.5 × 10-5 mol L-1 and total time of analyses of 20 s yielding an analytical frequency of 195 h-1 with a low waste generation per analysis (480 µL).

Palabras clave

Low temperature, Co-fired ceramic, Analytical microsystems, Dipyrone, Flow injection, LED photometer

Citación