A respirometry system designed for small ruminants.
| dc.contributor | UNIVERSIDADE FEDERAL RURAL DE PERNAMBUCO | |
| dc.contributor | UNIVERSIDADE FEDERAL RURAL DO RIO DE JANEIRO | |
| dc.contributor | UNIVERSIDADE ESTADUAL JULIO DE MESQUITA FILHO | |
| dc.contributor | RAFAEL GONCALVES TONUCCI, CNPGL | |
| dc.contributor | THIERRY RIBEIRO TOMICH, CNPGL | |
| dc.contributor | FERNANDA SAMARINI MACHADO, CNPGL | |
| dc.contributor | MARIANA MAGALHAES CAMPOS, CNPGL. | |
| dc.creator | SILVA, C. S. | |
| dc.creator | DIAVÃO, J. | |
| dc.creator | MOTTA, E. F. | |
| dc.creator | SILVA, A. S. | |
| dc.creator | TONUCCI, R. G. | |
| dc.creator | TOMICH, T. R. | |
| dc.creator | MACHADO, F. S. | |
| dc.creator | CAMPOS, M. M. | |
| dc.date | 2025-08-28T15:48:36Z | |
| dc.date | 2025-08-28T15:48:36Z | |
| dc.date | 2025-08-28 | |
| dc.date | 2025 | |
| dc.date.accessioned | 2026-07-07T04:00:29Z | |
| dc.description | This technical note describes a small ruminant respiration chamber system designed to accurately quantify the production of carbon dioxide (CO2) and methane (CH4). The system consists of 3 open-circuit respiration chambers, flow meters, gas analyzers, and an accessible environmental control system. To validate its performance, gas recovery tests were conducted by injecting CO2 and CH4 at 4 constant flow rates: 0.4, 0.8, 1.2, and 1.6 standard liters per minute (SLPM) for CO2, and 0.16, 0.20, 0.24, and 0.28 SLPM for CH4. Each injection level was tested across the 3 chambers. The average percentage recoveries for CO2 and CH4 were 97.8% ± 1.46% for chamber 1, 101.1% ± 1.49% for chamber 2, and 104.1% ± 1.98% for chamber 3 and 98.7% ± 0.97% for chamber 1, 99.4% ± 2.06% for chamber 2, and 101.6% ± 2.90% for chamber 3, respectively, demonstrating the system’s accuracy in quantifying livestock GHG emissions from small ruminants, such as young cattle, goats, and sheep. Therefore, this system represents a viable methodology for nutritional and environmental studies with small ruminants. Future studies conducted in this facility can contribute to a better understanding of the dynamic and flow of GHG emissions in small ruminants and target measures to increase the sustainability of small ruminant production systems. | |
| dc.identifier | JDS Communications, v. 6, p. 669–674, 2025. | |
| dc.identifier | http://www.alice.cnptia.embrapa.br/alice/handle/doc/1178364 | |
| dc.identifier | https://doi.org/10.3168/jdsc.2025-0755 | |
| dc.identifier.uri | http://hdl.handle.net/123456789/450419 | |
| dc.language | eng | |
| dc.rights | openAccess | |
| dc.subject | Câmara respiratória | |
| dc.subject | Bovino | |
| dc.subject | Ruminante | |
| dc.title | A respirometry system designed for small ruminants. | |
| dc.type | Artigo de periódico |