X-ray and gamma-ray computerized minitomograph scanner for soil science.

dc.contributorEMBRAPA-CNPDIA; Universita di Roma "La Sapienza", Centro D'Ingigneria Biomedica.
dc.creatorCRUVINEL, P. E.
dc.creatorCESAREO, R.
dc.creatorCRESTANA, S.
dc.creatorMASCARENHAS, S.
dc.date2025-06-27T19:48:12Z
dc.date2025-06-27T19:48:12Z
dc.date1999-12-22
dc.date1990
dc.date.accessioned2026-07-07T04:29:50Z
dc.descriptionAbstract-A computerized tomograph scanner system using X- and y-rays has been developed for applications in soil science. Previous results were obtained using a miniature X-ray tomograph scanner designed for biomedical analysis [ l]. As a new methodology of instrumentation and in soil research [2], this apparatus has proven to be useful for measuring volumetric water content 8 to an accuracy of *3% and soil bulk density p f 2% (in grams/centimeters3 ). The system features translation and rotation scanning modes, a 200- mm effective field of view, signal processing by pulse counting and 1.0- mm spatial resolution. The performance of the system has been demonstrated by experimentally measuring water content and the bulk density of soil samples.
dc.identifierIEEE Transactions on Instrumentation and Measurements, New York, v.39, n.5, p.745-750, Oct. 1990.
dc.identifier1557-9662
dc.identifierhttp://www.alice.cnptia.embrapa.br/alice/handle/doc/26404
dc.identifier10.1109/19.58619
dc.identifier.urihttp://hdl.handle.net/123456789/463181
dc.languageeng
dc.rightsopenAccess
dc.subjectstructure and physiological function
dc.subjectmedicine
dc.subjecttomography
dc.titleX-ray and gamma-ray computerized minitomograph scanner for soil science.
dc.typeArtigo de periódico

Archivos